Kalibrasi CT Scan Merk Siemens Type Somatom go.Top di RSUD Banyumas

Authors

  • Agustinus Lise Sekolah Tinggi Ilmu Kesehatan Semarang
  • Diah Rahayu Ningtias Sekolah Tinggi Ilmu Kesehatan Semarang
  • Imam Tri Harsoyo Sekolah Tinggi Ilmu Kesehatan Semarang

DOI:

https://doi.org/10.55606/jurritek.v3i2.7469

Keywords:

CT Scan, Calibration, Uncertainty, Imaging, X-ray

Abstract

CT Scan calibration is necessary to ensure the accuracy, consistency, and reliability of the medical imaging results used in the diagnosis process. This study aims to find out the calibration procedure and analyze the data of the calibration results of the Siemens SOMATOM go CT Scan. TOP at Banyumas Hospital. The test method refers to the Ministry of Health's CT Scan Testing Working Method No. MK: 056-18, which is the standard in evaluating the performance of CT Scan equipment. The parameters tested included the percentage of tube voltage error (kVp) and the quality of the X-ray beam represented by the Half Value Layer (HVL) value. The results of the measurement of the error percentage at various kVp points show a range of 0.37–0.88%, which is still within the tolerance limit set by the national standard. In addition, the results of the HVL test at voltages of 120 kVp and 140 kVp also showed values that were above the minimum required limit, so that the quality of the X-ray beam was considered to meet the requirements. Based on the overall test and analysis results, the Siemens SOMATOM CT Scan go. TOP is declared feasible, safe, and good to use because all performance parameters of the tool have met applicable standards.

Downloads

Download data is not yet available.

References

N. K., & I. S. (2010). Evaluasi nilai noise sebelum dan sesudah kalibrasi CT scan. ugm.ac.id.

Ary Kurniawati, Cholid Badri, & Jeffri Ardiyanto. (2023). Desain phantom air untuk kalibrasi CT number dan keseragaman CT number pada pesawat CT Scan. Jurnal Imejing Diagnostik (JImeD).

Ayu, R. G. (2018). Analisis noise berdasarkan slice thickness dengan teknik irisan axial pada citra CT Scan. unej.ac.id.

Ayu, R. G. (2018). Analisis noise berdasarkan slice thickness dengan teknik irisan axial pada citra computed tomography scan (CT-Scan) (Skripsi). Jurusan Fisika, Universitas Jember.

Fransiska, Aditya. (2024). Uji kesesuaian berkas sinar-X mobile di Rumah Sakit Bhayangkara Jambi (Skripsi). Program Studi Fisika, Universitas Jambi.

Hussain, S., et al. (2022). Modern diagnostic imaging technique applications and risk factors in the medical field: A review. PMC Articles. https://doi.org/10.1155/2022/5164970

Ilmu, R. (2018). Dosimeter. rakusilmu.home.blog.

International Atomic Energy Agency. (n.d.). Handbook of basic quality control tests for diagnostic radiology. IAEA.

Kesehatan, U. (2009). Undang-undang Republik Indonesia nomor 36 tahun 2009 tentang kesehatan.

Mullen, L. J. (2020). Understanding X-ray images: A medi(c)a(l) aesthetics approach. Dalam S. Josephson, J. D. Kelly, & K. Smith (Eds.), Handbook of visual communication: Theory, methods, and media (Edisi ke-2, pp. 304–320). Routledge. https://doi.org/10.4324/9780429491115-28

Nariswari, N. N. (2018). Analisa variasi faktor eksposisi dan ketebalan irisan terhadap CT Scan. unej.ac.id.

Permenkes. (2015). Peraturan Menteri Kesehatan Republik Indonesia nomor 54 tahun 2015. Kementerian Kesehatan RI.

Rapih, R. S. (2021). Pemindaian tomografi terkomputasi (CT Scan). pantirapih.or.id.

Schiffers, F., Bochynek, T., Aichert, A., Würfl, T., Rubenstein, M., & Cossairt, O. (2020). Disassemblable fieldwork CT scanner using a 3D-printed calibration phantom. arXiv.

Sutanto, A. R. (2014). Uji kesesuaian CT number pada pesawat CT Scan multi slice.

Downloads

Published

2024-10-30

How to Cite

Agustinus Lise, Diah Rahayu Ningtias, & Imam Tri Harsoyo. (2024). Kalibrasi CT Scan Merk Siemens Type Somatom go.Top di RSUD Banyumas. JURAL RISET RUMPUN ILMU TEKNIK, 3(2), 96–106. https://doi.org/10.55606/jurritek.v3i2.7469

Similar Articles

You may also start an advanced similarity search for this article.