Analisis Pengaruh Variasi Slice Thickness Terhadap Kualitas Citra Pemeriksaan CT Scan Urografi Pada Kasus Nefrolitiasis Di RS TK. II Pelamonia Makassar
DOI:
https://doi.org/10.55606/jikg.v2i1.2147Keywords:
Nefrolitiasis, CT Scan Urografi, Slice thicknessAbstract
Background: Kidney stones (nephrolithiasis) are solid masses formed from urine precipitated crystals. The formation of stones is caused by increased calcium, oxalate and uric acid content in the body, scans to view kidney stones (nephrolithiasis) can be performed with or without contrast. Non-contrast CT urography is the imaging modality of choice for the initial evaluation of patients with suspected urolithiasis, as it has a sensitivity of up to 98% and specificity of 96-98%. Components that affect image quality in CT scans are spatial resolution, contrast resolution, and noise. One of the important parameters in image quality is the selection of slice thickness. Spatial resolution is the ability to show objects or organs with a high level of contrast. Noise is the standard deviation percentage of the sum of pixel values obtained from the waterbath image.
Methods; This type of research is quantitative research with an experimental approach to analyze the effect of slice thickness variations on image quality using variations of 2 mm, 3 mm, 4 mm, 5 mm, and 6 m. The data were collected in August 2023 with 10 samples. Data were collected in August 2023 with 10 samples. Researchers took urographic CT scan examination data with a diagnosis of nephrolithiasis.
Results and Conclusion: With the results of the research conducted, it can be concluded that in the overall Friedman test results there is an effect of using slice thickness variations on image quality in urographic CT Scan examinations with nephrolithiasis diagnoses. using slice thickness variations. Based on the results of this study, the researcher can recommend on CT Scan urography with nephrolithiasis diagnosis using 4 mm slice thickness to get good and optimal image quality results.
References
Hofer M. CT teaching manual system approach to CT reading. 4th edition. New York: Springer-Verlag; 2000.
Moore, R., Harvin, H. dan Chow, L. (2018) ‘Pitfalls in MDCT urography’, AppliedRadiology,47(12),pp.16 21.Available at https://appliedradiology .com/communities/CT-Imaging/pitfalls-in-mdCT- urography
National Institute of Diabetes and Digestive and Kidney Diseases. (2015). BladderStones.https://www.niddk.nih.gov/health-information/urologic- diseases/bladder-stones.
Kementerian Kesehatan, R.I., 2013. Riset Kesehatan Dasar (Riskesdas) 2013. Jakarta: Badan Penelitian dan Pengembangan Kesehatan
Pramiadi, P., Utomo, B. P. dan Setyawan, N. H. (2017) ‘Dual-Energy Computed Tomography untuk Menentukan Komposisi Batu Urin’, Jurnal Radiologi Indonesia, 2(2), pp. 104–116. doi: 10.33748/jradidn.v2i2.54.
Nugroho RA, Ardiyanto J, Wijokongko S. Analisis Variasi Slice thickness terhadap informasi anatomi potongan axial pada pemeriksaan MSCT Cervical pada kasus trauma. J Imejing Diagnostik. 2020;6(2):91-5.10.
Listiyani II, Nismayanti A, Maskur M, Kasman K, Ulum MS, Rahman AR. Analisis Noise Level Hasil Citra CT-Scan Pada Phantom Dengan Variasi Tegangan Tabung Dan Ketebalan Irisan Garvitasi. 2021;20(1)5-9.
Dale dkk (2013) .Imaging Advances in Nefrolitiasis. Journal of Endourology, 31(7)’, Imaging Advances in Nefrolitiasis. Journal of Endourology,31(7), (September), pp.1– 18.doi:10.1089/akhir.2016.0695
M akmur, I Wayan Ari; Setiabudi, Wahyu; Anam C. Evaluasi Ketebalan Irisan (Slice Thickness) pada Pesawat CT-Scan Single Slice [Internet]. Vol. 21, Jurnal Sains dan Matematika. 2015. p. 6. Available from: https://ejournal.undip.ac.id/index.php/sm/article/view/8032
Almuslimiati A, Milvita D, Prasetio H. Analisis Nilai Noise dari Citra Pesawat CT-Scan pada Beberapa Rekonstruksi Kernel dengan Variasi Slice Thickness. J Fis Unand [Internet]. 2019;8(1):57–63. Available from:http://jfu.fmipa.unand.ac.id/index.php/jfu/article/view/392/35
Downloads